| Frequency Generation |
|
| Frequency Range |
9 KHz to 3 GHz |
| Frequency Resolution |
0.1 Hz |
|
|
| Frequency Stability |
|
| Ageing Rate for reference frequency |
<2 X 10-6 / year |
| Temperature stability for reference frequency |
<2 X 10-6 |
|
|
| Spectral purity |
|
| SSB phase noise at 1 GHz, on 20 KHz offset and 1 Hz bandwidth |
<-95 dBc / Hz |
Harmonics at 3 GHz at levels 0 dBm
|
< -30 dBc |
| Non harmonics (>10 KHz carrier offset) |
< -50 dBc |
| |
|
| Reverse power protection |
2 W |
| |
|
| Output Level |
|
| Range |
-125 dBm to +13 dBm |
| Resolution |
0.1 dB |
| Total accuracy (for levels > -120 dBm) at 25 Deg C |
<± 1 dB |
| VSWR |
<1.5 at 1 GHz and <2 at other frequencies. |
|
|
| Modulations |
AM, FM or Phase modulation,
IQ and Pulse modulation |
| Amplitude Modulation |
Internal, External |
|
|
| Modulation freq. Range (3 dB), f > 100 KHz |
DC/ 20 Hz to 20 KHz |
| Modulation Depth |
0 to 99.99% |
| Resolution |
0.1% |
| AM accuracy at 1 KHz (m < 80%) |
< 5% of reading |
|
|
| Frequency Modulation |
Internal, External |
| Modulation freq. Range (-3 dB) standard |
DC/20 Hz to 80 KHz |
| Frequency Deviation |
1 MHz for 512 MHz < fc < 3 GHz
500 KHz for 256 MHz < fc < 512 MHz
250 KHz for 128 MHz < fc < 256 MHz
125 KHz for 200 KHz < fc < 128 MHz
0.1fc for fc < 200 KHz
|
| Resolution |
1% min. 1 Hz |
| Setting Error (at AF=1 KHz and 50% max deviation) |
<5% of reading + 300 Hz
|
| |
|
| Phase Modulation |
Internal, External |
| Phase Deviation |
0 to 5 rad (min.) at 3 GHz carrier. |
| Resolution |
0.01 rad (1%) |
| Setting Error(at AF = 1 KHz) |
<5% of reading + 0.2 rad |
|
|
| Pulse modulation with built in Pulse generator |
| Rise /Fall time |
<3 micro sec |
|
|
| I/Q Modulation |
External DC to 40 MHz |
|
|
Internal Modulation Generator
(for AM/FM/Phase M) |
20 Hz to 80 KHz with resolution of 0.1 Hz |
| |
|
| General features: |
|
| Memory |
100 storage settings |
| Digital Sweep |
RF, LF in continuous sweep, single sweep, and single step . |
| Interface |
USB, RS232 , IEEE-488 |
| Display |
16 digit x 2 line LCD |
| Operating temperature Range |
0 to 55°C |
| Power Requirement |
230 V ± 10% AC, 50 Hz ± 3% |